Application
The coolers have the function of lowering the gas dew point to prevent condensation in the analyzer.
A good and stable dew point to prevent cross-sensitivity among different H 2 O
The dew point is adjustable between 1 ° C and 15 ° C.
The digital control has several options and outputs, such as mA output, RS 485 and Modbus communication with digital inputs.
This Peltier gas cooler have extra powerful peltier elements and instead of a traditional aluminum heat sink, a superconducting forced heat, so it can deliver much more power than the conventional Peltier coolers.
 
Description
This cooler has recently entered the market and has incorporated all the innovations and techniques that we need for a highly accurate, precise, safe and sustainable cooler to ensure stable measurement. The unique cooling principle with our patented technology ensures a low and constant dew point of 4 ° Celsius, which for a stable and reliable measurement. The cooler can be easily adapted to various types of heat exchangers
 
The cooler compensates for variations in the gasflow and varying ambient temperatures.
 
The extremely low resolution is obtained by the new cooler and the patented heat exchanger technology. The design of the heat exchanger for a so-called shock effect of the cooling gas  minimises the contact time between the condensate and the gas. Therefore low SO2 standard can be measured without  additives and the like.
 
The heat exchangers are easy to exchanges and to clean. 
 
The cooler is electronic controlled. The signals can be connected to an electronic console, such as our ADSs, which the other gasconditioneringscomponenten can be read.
 
Obviously, the cooler is also equipped with conventional NO alarm contacts.  

  • Special removable heat exchangers, which also can be easily developed
  • Humidified heat exchangers for calibration to eliminate cross-sensitivity
  • Digitally controlled allowing a very stable dew point arises. ± 0.1 ° C
  • Ambient to 45 ° C!!
  • Alarm contact
  • Optional digital communications Modbus / RS485
  • Power 115 to 230 VAC

 

Model APC
APC 4xx
APC 5xx
APC 6xx
Aantal warmtewisselaars
1
2
1
Behuizing
RVS, wandmontage
Afmetingen (HxLxD)
180 x 280 x 180mm
 
 
Data per warmtewisselaar
APC 4x1/5x1/6x1
APC 4x2/5x2/6x2
APC 4x3/5x3/6x3
Maximale doorstroming
1x 150Nl/h max.
2x 150Nl/h max.
1x 350Nl/h max.
Materiaal warmtewisselaar
Duran® Glas, PTFE
PVDF, PTFE
SS316
Maximum druk
3 bar a
3 bar a
10 bar a
Dood volume
35 cm³
35 cm³
100 cm³
Sample gas ingang
1x G1/4”i
1x G1/4”i
1x G1/4”i
Sample gas uitgang
1x G1/4”i
1x G1/4”i
2x G1/4”i
Condensaat uitgang
1x GL25
1x 3/8”i
1x G3/8”i
 
Operation data
APC 4xx
APC 5xx
APC 6xx
Gas ingang dauwpunt
Max. 65°C*
Gas ingang temperatuur
Max. 190°C*
Gas uitgang dauwpunt
+1°C ..... +15°C, fabrieksinstelling: +4°C
Stabiliteit
0,1°C bij 20°C omgevingstemperatuur
Omgevingstemperatuur
+5°C  tot 45°C
 
Electrical data general
APC 4xx
APC 5xx
APC 6xx
 
Alarm contact
Vrij programmeerbaar contact 1NO / 1NC, vermogen: 250V, 16A AC
Beschermingsklasse
IP20 EN 60529 / EN 61010
Vermogen:
75W
 
Model APC
APC 40x
APC 50x
APC 60x
Voeding
230VAC, 50/60Hz
 
 
Model APC
APC 41x
APC 51x
APC 61x
Voeding
115VAC, 50/60Hz
Maximum values in technical data’s must be rated in consideration of total cooling capacity at 25°C ambient temperature and 4°C outlet dew point
PTFE               = Polytetrafluoroethylene (Teflon®)
PVDF            = Polyvinylidenfluoride

Documentatie