OPERATING
PRINCIPLE
The SO3 or H2SO4 in the gas sample is
absorbed as sulphate ions (SO42-) in an aqueous solution of propan-2-ol in
water.
The solution is passed through a bed
of barium chloranilate. The acid chloranilate ions created are measured in a
continuous
flow photometer. By maintaining a
constant ratio of flow rates for the gas and the propan-2-ol absorbing solution,
the concentration of acid chloranilate
ions can be directly related to the sulphate ion concentration in the absorbing
solution,
and hence the SO3 concentration in the
gas.
TECHNICAL SPECIFICATION
Range 0 - 200 ppm
Accuracy +/- 5% of reading (in
calibrated range)
Lag time Approx. 60 seconds
Response (90% FSD) Approx. 120 seconds
(depending on umbilical length)
Solution consumption 0.25 - 2 cm3/h
(adjustable as required)
Ambient temperature 0 - 40?C
Module dimensions
Analysis 19? unit, 6U high, 500mm deep
(27 kg)
Control 19? unit, 4U high, 500mm deep
(13 kg)
Power Requirement 230V 50/60 Hz or
110V 50/60 Hz, 350 W
Probe Length 1 m, 1.2 m, 1.5 m, 2 m
Signal Output 4 .. 20 mA output for
Recorder
RS232 to PC connector on request